Checks
WARNING
MR system malfunction!
Hazardous conditions for patients
◆ Note the sounding alarm and signal.
◆ Do not perform MR examinations.
◆ Notify Siemens Service.
(1) LINE POWER LED
(2) HELIUM LEVEL LED
(3) MAG STOP LED
(4) PRESSURE LED
(5) COMPRESSOR LED
(6) BATTERY LOW LED
(7) EIS LED
(8) COMMS FAULT LED
(9) ACKNOWLEDGE button
(10) Magnet Stop button
LED LEDs light up to indicate
LINE POWER Voltage supply of MR system is satisfactory
HELIUM LEVEL Helium fill level is too low
MAG STOP Magnet Stop switch is pressed / quench
triggered
or:
Error in the Magnet Stop circuit
3.6.2
Checking the LEDs (version 1)
MR system components 3
MAGNETOM Avanto, MAGNETOM Verio | syngo MR E11
67