Siemens AG SPB7-230.841.03 Page 1 of 2 M300/1000/3000/3000 Nova
Medical Solutions Rev. 02 04.04 CS PS 24 Service
Prerequisites 1
General 1
This instruction describes:
in Chapter 2 - how to replace the compression mechanism in the compression unit of the
mammography systems MAMMOMAT 300 and MAMMOMAT 1000/3000/3000 Nova.
in Chapter 3 - how to replace the potentiometer R863 - the position transducer of the
compression unit.
Components concerned 1
Components concerned are the compression units/devices normally used on
MAMMOMAT 1000/3000/3000 Nova systems, on MAMMOMAT 300 systems and - with
some modifications - also used on MAMMOMAT Novation DR systems.
Material numbers
063 86 051 (used for the Sales Item)
065 21 681 (used for the Sales Item)
064 31 725 (used as Spare Part with REP-Status)
If you encounter compression units with material numbers that are not listed here, apply
this instruction appropriately. If necessary, contact CS HSC 24.
Tools required 1
Standard toolkit (including Allen keys with ball-shaped heads).
Resistance Measuring Instrument (e.g. Fluke - with Ohm-meter function).
Documents required 1
The technical documentation delivered with the corresponding MAMMOMAT system.
Mainly: Maintenance Instruction SPB7-230.105...
Service Program SPB7-230.114...
Strictly observe all safety precautions - as described in the
relevant technical documents - during work that involves
assembly, disassembly, removal of covers or work under voltage.
NOTICE