Loading...
2 - 1
Siemens AG SPB7-230.841.03 Page 1 of 6 M300/1000/3000/3000 Nova
Medical Solutions Rev. 02 04.04 CS PS 24 Service
Procedure - Compression mechanism 2
Remove the covers from compression unit 2
System OFF.
Pry open the lid on the maximum compression force presetting knob. Loosen the screw
underneath the lid and remove the knob (1/Fig. 1).
Remove the cover from the manual compression/decompression knobs (2/Fig. 1).
Release the locking pin on the upper and lower cover and remove the covers.
Loosen the Allen screws on top and at the bottom of the compression unit holding the left
side cover (3/Fig. 1). Remove the side cover.
Fig. 1
3
3
Locking pin
1
2