Procedure - Compression mechanism 2 - 3
Siemens AG SPB7-230.841.03 Page 3 of 6 M300/1000/3000/3000 Nova
Medical Solutions Rev. 02 04.04 CS PS 24 Service
• Loosen the Allen screws holding the plate that keeps the curtain in place (1/Fig. 3).
Remove the plate and the curtain.
• Loosen the Allen screw securing the spindle (2/Fig. 3).
• Push the spindle inwards until it reaches the back of the compression unit in order to
release the compression mechanism.
Fig. 3
Ensure that the manual compression/decompression knobs are
not in their lowest position. In the lowest position it is not
possible to push the spindle fully in towards the back of the
compression unit and the compression mechanism will not be
released.
Curtain
Spindle
1
2
NOTE