Siemens Elema AB Register 6 SPB7-230.051.07 Page 15 of 16 M1000/3000 Nova
Solna, Sweden Rev. 02 03.00 SPS-UD Wiring Diagrams
X041E 2-15
LIST OF VOLTAGES AND TEST POINTS, STAND
Voltage Description ( typical value)
+5V Regulated DC voltage, 4.75V < +5V < 5.25V
+5V_REF DC voltage, +5V -5% < +5V_REF < +5V
+5V_DSP Regulated DC voltage, 4.75V < +5V_DSP < 5.25V
+7V Regulated DC voltage, 6.60V<+7V<7.25V
+5V_MPS Regulated DC voltage, 4.75V < +5V_MPS < 5.25V
0VD Digital ground
0VA Analog ground
0V_MPS Generator interface ground. Floating, grounded by generator.
11V AC supply voltage, 11VAC
1) 2)
11V_MPS AC supply voltage, 11VAC
1) 2)
+5V Regulated DC voltage, 4.75V < +5V < 5.25V
+5V_M Regulated DC voltage, 4.75V < +5V_M < 5.25V for optointerface D801-D805
0V_W Digital ground optointerface D801 - D805
0V_M Digital ground optointerface D801 - D802
+14V Unregulated DC voltage, 18V
1)
+15V Regulated DC voltage, 14.25V < +15V < 15.75V
+24V Unregulated DC voltage, 32V
1)
+24VF Unregulated DC voltage, 32V
1)
14VAC AC supply voltage, 14 VAC
1) 2)
17VAC AC supply voltage, 17 VAC
1) 2)
17VACF AC supply for lamp
1)2)
0V_L Floating ground for LAMP voltage regulation
1)2)
-V_L Unregulated supply voltage for lamp control circuit
1)2)
24VAC AC supply voltage, 24VAC
1) 2)
24VACF AC supply voltage for the fan
1)
2)
0V Circuit ground
+5V Regulated DC voltage, 4.75V < +5V < 5.25V
+5V_R Regulated DC voltage, 4.40V < +5V_R < 5.25V