Siemens Elema AB Register 6 SPB7-230.051.07 Page 16 of 16 M1000/3000 Nova
Solna, Sweden Rev. 02 03.00 SPS-UD Wiring Diagrams
2-16 X041E
1)
STANDBY TYPICAL VALUE, 230VAC INPUT
2)
Floating voltage. Measure with digital multimeter or differential amplifier.
Page Voltage Origin Supplied
from
Supply for Test LED Fuse
3-3/2B 230VAC Generator D711/X14 Stand X881/2,3
3-3/1C P.E. Stand X899 Inc. mains voltage Whole system X899/PE
3-3/1C N Stand X899 Inc. mains voltage Whole system X899/N
3-3/1C L1 Stand X899 Inc. mains voltage Whole system X899/L1
3-3/1C L2 Stand X899 Inc. mains voltage Whole system X899/L2
3-3/1C L3 Stand X899 Not used X899/L3
Voltage Description ( typical value)
230VAC Stand mains voltage input, 207VAC < 230VAC < 253VAC
P.E. Protective earth terminal, incoming mains
N Neutral line terminal, incoming mains
L1 Line 1 terminal, incoming mains voltage
L2 Line 2 terminal, incoming mains voltage
L3 Not used
LIST OF VOLTAGES AND TEST POINTS, STAND