Siemens Elema AB Register 6 SPB7-230.051.07 Page 3 of 14 M1000/3000 Nova
Solna, Sweden Rev. 02 03.00 SPS-UD Wiring Diagrams
X041E 4-3
CONTROL PANEL FUNCTIONS
Fuses 0
NOTE! See spare parts list for replacement fuses.
1)
See note on page 4-1.
LOCATION FUSE FUSED VOLTAGE
Incoming mains F10 20 AT (35 AT) /500 V L1
1)
Incoming mains F20 20 AT/500 V L2
1)
D711 F1 Not used
D711 F2 4 AT/250 V 230V_T1
1)
D711 F3 4 AT/250 V 240V_T1
1)
D711 F4 4 AT/250 V 230V_STAND
D711 F5 4 AT/250 V N_T1
1)
D711 F6 1.5 AT/250 V PRIM1_T2
D711 F7 1 AT/250 V 19V_T1
D711 F8 1.5 AT/250 V UZ
D711 F9 1.5 AT/250 V N_UZW
T2 F3 6.25 AT/250 V U1_HEIZ
T2 F4 0.5 AT/250 V U2_24V
T2 F5 1 AT/250 V U2_13V
T2 F6 1.5 AT/250 V U2_+5V
T2 F7 1.5 AT/250 V U2_+15V
T2 F8 1.5 AT/250 V U2_-15V
T3 F1 1 AT/250 V 24V_AC1