Collimator lamp replacement
MAMMOMAT Balance SPB7-115.840.01 Page 10 of 62 Siemens AG
Troubleshooting Guide Rev.02 10.04 Medical Solutions
Collimator lamp replacement
1) Remove attachment of Face shield protection by means of four fixing screws
2) Open collimator assembly by removing the four fixing screws
3) replace lamp only with original spare part having the same filament orientation.