Rotating anode starter
MAMMOMAT Balance SPB7-115.840.01 Page 25 of 62 Siemens AG
Troubleshooting Guide Rev.02 10.04 Medical Solutions
Function configuration
Default Jumper setting
JP1 = open Brake enabled
JP2 = open not used
JP3 = shorted Brake time 6”
JP4 = open Protection enabled
JP5 = refer to above image
Brake disable for heavy use
Anode rotation brake is not essential for unit operation and functionality.
Brake is factory enabled, to avoid significant tube temperature increase it’s recommended to disable
brake ( JP1 Open ) for screening or equivalent heavy duty.
Brake energy can be equivalent to Rotation start + average exposure.