3Replacing basic unit electrical assemblies
Replacing the PXCM (BRICK) 0
General 0
These instructions describe how to disassemble the PXCM (BRICK) component and
mount a new one. It then describes how to perform subsequent required tasks, mainly
checks of several adjustments.
If a BRICK has to be replaced, the spare part material number 066 46 728 (with the
REP-status) must be ordered.
When describing various tasks, these instructions address the differences between the
two workstations (WH AWS and DROC).
Documents required 0
The type of documentation depends on the acquisition workstation in use:
• DROC
- Installation Instructions and Startup (DROC)
- Quality Control manual (DROC)
• WH AWS
- Startup, WH AWS VAXXX (XXX - Depends on the installed software version)
- Software, WH AWS VAXXX (XXX - Depends on the installed software version)
- Quality Control manual (WH AWS)
Tools and test equipment required 0
• Drill with a 5.5 mm drill bit
• Dosimeter for mammography
• Power ground-wire tester
• 24 x 30 compression plate
• 3x 20 mm plexi (PMMA) phantom
• Plexi phantom 42x350x350mm
• Alu sheets 1x0.1 + 2x0.2 + 1x0.5 (8881273)
Establishing a connection to the BRICK 0
There are two software interfaces available on the BRICK, the web-based interface and
the command line interface.
Establishing a connection to the BRICK web interface
To access the web-based BRICK interface, start a web browser on the acquisition work-
station and enter the appropriate URL.
The procedures differ for opening the BRICK Web home page on a syngo-based acquisi-
tion workstation (WH AWS) and a DROC acquisition workstation.