Siemens AG SPB7-250.841.01.04.02 MAMMOMAT Novation DR
05.05 CS SD 24
Replacing Stand Boards 39
Page 39 of 64
Medical Solutions
D801 stand CPU board 0
Fig. 14: D801
• Make a backup of the stand data. See Software System Stand and Backup/Restore.
• Turn the system OFF.
• Remove the right stand cover.
• Move prom J9 and J65 from the old to the new Board (1/Fig. 14 / p. 39).
• Install the new board.
• Install the right stand cover.
• Turn the system ON.
• Restore the stand data.