44 Collimator and D814
MAMMOMAT Novation DR SPB7-250.841.01.04.02 Siemens AG
05.05 CS SD 24
Page 44 of 64
Medical Solutions
Disassembling board D814 0
Fig. 20: Collimator
Fig. 21: D814
Pos. 1 X838
Pos. 2 X837
Pos. 3 X836
Pos. 4 X835
Pos. 5 X834
Pos. 6 X833
• Remove the shielding plate from board D184 (1/Fig. 20 / p. 44).
• Loosen connectors X833, X834, X 835, X836 and X838 (Fig. 21 / p. 44).
• Loosen board D814 from the 4 plastic holders (1/Fig. 22 / p. 45).
• Board D814 can be removed from the collimator.