Siemens AG SPB7-250.841.01.04.02 MAMMOMAT Novation DR
05.05 CS SD 24
Replacing AWS Celsius 420 63
Page 63 of 64
Medical Solutions
• Pay attention to the fiber optic cables (connections between Brick and PCI card). These
cables and plugs should be handled with care (1/Fig. 39 / p. 62).
Fig. 40: AWS
• Remove the dongle from the old computer and attach it to the new one
(1/Fig. 40 / p. 63).
• Connect the connecting cables to the new computer.
• Software installation:
The "WH AWS Software Reinstallation" chapter in the technical documentation
"Software, System WH AWS, SPB7-250.816...." (depending on installed WH AWS
version) describes the steps necessary for installing the software.
You can refer to the "Start-up, WH AWS(SPB7-250.815.02..)" if not all settings have
been backed up and some settings are missing after the restore operation.
Final steps 0
Perform a test exposure with a phantom.