Siemens AG SPB7-250.841.01.04.02 MAMMOMAT Novation DR
05.05 CS SD 24
Replacing AWS Celsius 420 63
Page 63 of 64
Medical Solutions
Pay attention to the fiber optic cables (connections between Brick and PCI card). These
cables and plugs should be handled with care (1/Fig. 39 / p. 62).
Fig. 40: AWS
Remove the dongle from the old computer and attach it to the new one
(1/Fig. 40 / p. 63).
Connect the connecting cables to the new computer.
Software installation:
The "WH AWS Software Reinstallation" chapter in the technical documentation
"Software, System WH AWS, SPB7-250.816...." (depending on installed WH AWS
version) describes the steps necessary for installing the software.
You can refer to the "Start-up, WH AWS(SPB7-250.815.02..)" if not all settings have
been backed up and some settings are missing after the restore operation.
Final steps 0
Perform a test exposure with a phantom.