3 - 1
Siemens AG SPB7-230.814.01 Page 1 of 2 Opdima
Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Protective measures 3
It is very important that any intervention in the equipment shall start with disconnecting it
from the power supply with the main circuit breaker. To prevent accidental triggering of
high voltage and radiation, set the switch S2 (SS) on board D750 to OFF.
When switching off the workstation use the power off procedure
described in the Supplement to the Instructions for Use MAMMO-
MAT Novation
DR
- Opdimar.
Switching off the workstation before the software has been
closed down may cause damage to the files on the hard disc.
If the system is only switched off at the control panel or with the
circuit breaker CB S700, there is still voltage present within the
generator. See the wiring diagram for MAMMOMAT Novation
DR
.
Life-threatening electric shock hazard exists.
Disconnect mains cable and comply with the information on this
page.
After shut-down of the system, there may still be about 380 V DC
present on the intermediate circuit of the MAMMOMAT generator.
Life-threatening electric shock hazard exists.
A bleader resistor is connected over the output DC capacitors.
The DC voltage will discharge below 60 Volt in 3 minutes.
The voltage level is indicated by LEDs H3 and H4.
Observe the currently valid guidelines for handling electronics
endangered by electrostatic discharge.
Use ESD-equipment, ground prior to making contact and place
the components on a conductive surface.
The boards contain electrostatic highly sensitive components
requiring particular care in their handling.
Risk of damaging components.
CAUTION
WARNING
WARNING
CAUTION