Opdima SPB7-230.814.01 Page 6 of 24 Siemens AG
Installation Rev. 01 10.04 CS PS 24 Medical Solutions
5 - 6 Installation
5
Cables to CCD camera 5
The CCD camera including cable with connectors marked X201 and X202 is connected to
the biopsy controller.
1. Connect the cable to X201 and X202 on the biopsy controller, see Fig. 4.
2. Remove the screws (1/Fig. 5) and remove the cover (2/Fig. 5) of the object table
arm.
Sensitive electronic equipment!
Risk of damaging the camera.
Do not touch the pins in the contacts of the CCD camera.
OPD00453
X201
X202
X203
X204
Ferrite
X205 X206 X207 Mains cable
Protective earth cable
Fig. 4 Connection of cables to the biopsy controller
CAUTION
OPD00161
1 2
Fig. 5 Object table arm before installation