Installation 5 - 9
Siemens AG SPB7-230.814.01 Page 9 of 24 Opdima
Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Installation of workstation 5
Preparing the workstation table (option) 5
The workstation table (option), a motorized vertically adjustable table, shall be used to put
the workstation components to the Opdimar system on. The vertical adjustment device
includes a transformer, which is not mounted on delivery.
1. Remove the screws on the underside of the tabletop and use them when
mounting the transformer (1/Fig. 9).
Place the transformer so that the connectors are turned to the column.
2. Connect the cable from the control unit (2/Fig. 9) for the vertically adjustment to
the connector marked “Handschalter” (3/Fig. 9).
3. Connect the cable from the motor (4/Fig. 9) for the vertically adjustment to the
other connector (5/Fig. 9).
The workstation table must not be overloaded.
Any additional equipment is added at your own risk and the maxi-
mum load of 80 kg must not be exceeded.
CAUTION
Opd00494
Handschalter
3
1
2
Fig. 9 Transformer
4
5