Opdima SPB7-230.814.01 Page 10 of 24 Siemens AG
Installation Rev. 01 10.04 CS PS 24 Medical Solutions
5 - 10 Installation
4. Place the workstation table outside the patient environment of safety reasons.
Patient environment is determined by IEC 60601-1-1 as 1.5 m away from the
patient.
There must be enough space around the workstation table so that
it safely can be vertically adjusted.
150
OPD00599
Fig. 10 System site overview
Cable duct
Stand
Biopsy
controller
CCD camera holder
CCD camera
Biopsy unit Generator
Cable (workstation table) 2 m
Workstation
M
i
n
.
1
,
5
m
Cable
Cables
8 m
1,5 m
NOTE