Opdima SPB7-230.814.01 Page 12 of 24 Siemens AG
Installation Rev. 01 10.04 CS PS 24 Medical Solutions
5 - 12 Installation
Connecting the workstation 5
1. Connect workstation cable marked X203 to the biopsy controller connector X203,
see Fig. 12, and the other end (X101) of the cable to the workstation connector
(2B/Fig. 26).
2. Attach ferrite on the cable on the biopsy controller side, see Fig. 12.
3. Connect the workstation cable marked X204 to the biopsy controller connector
X204 in Fig. 12 and the other end of the cable to the workstation connector
A (8/Fig. 17) and B (2A/Fig. 17).
4. Attach ferrite on the cable on the biopsy controller side, see Fig. 12.
5. Mount the touch protection over the connectors X203 and X204.
-Place the plastic plate (1/Fig. 13) behind the connectors X203 and X204.
-Place the plastic cover (2/Fig. 13) on the other side of the connectors.
-Fasten the plastic screws (3/Fig. 13).
Connecting the cable to the wrong connector may cause severe
damage to the equipment.
CAUTION
OPD00453
X201
X202
X203
X204
Ferrite
X205 X206 X207 Mains cable
Protective earth cable
Fig. 12 Biopsy controller
OPD0
Biopsy controller
1
2
3
Fig. 13 Mounting the touch protection