Prerequisites 1 - 3
Siemens AG SPB7-230.814.01 Page 3 of 4 Opdima
Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Documents required 1
• Supplement to the Instructions for Use MAMMOMAT Novation
DR
- Opdimar
(included in the Opdimar delivery)
• MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions
(included in the Opdimar delivery)
• MAMMOMAT 300/3000, MAMMOMAT 3000 Modular incl. Stereotactic Biopsy
Attachment and Opdimar DHHS Maintenance Instructions
(included in the Opdimar delivery)
• MAMMOMAT 300/3000, MAMMOMAT 3000 Modular incl. Stereotactic Biopsy
Attachment and Opdimar Measurement Certificates
(included in the Opdimar delivery)
• MAMMOMAT Novation
DR
- Opdimar Planning Guide
• MAMMOMAT Novation
DR
- Opdimar Installation and Start-Up
• MAMMOMAT Novation
DR
- Opdimar Wiring Diagram
• MAMMOMAT 3000 Installation and Setting Instructions Update Kit for Stereotactic
Biopsy Attachment (where applicable)
Meters and appliances required 1
• Protective ground wire tester (44 15 899 RV090)
• Service PC (e.g. Siemens-Nixdorf PCD 3NSX/20 or similar) with connecting cable, PC to
generator (part No. 99 00 440 RE999)
• Stereo calibration phantom (included in the delivery) (part No.64 30 701)
• AEC calibration plexiglass, four plates measuring 150 mm x 150 mm x 19 mm and one
plate measuring 150 mm x 150 mm x 9,7 mm, part No. 65 61 232 and 65 61 224
respectively
• Resolution phantom with at least 10 line pairs per mm
Tools required 1
• Standard installation tools
• Electric drill with 5 mm drill
Additional installation material required 1
• Extension mains cord with at least three outlets, of protective earth type
Start-up, DHHS and X-ray decree area of application 1
Notes on start-up 1
Opdimar is adjusted, programmed and tested in the factory, leaving only the connection
to the on-site mains voltage, the functional tests and customer dependent adjustments to
be performed.