Prerequisites 1 - 3
Siemens AG SPB7-230.814.01 Page 3 of 4 Opdima
Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Documents required 1
Supplement to the Instructions for Use MAMMOMAT Novation
DR
- Opdimar
(included in the Opdimar delivery)
MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions
(included in the Opdimar delivery)
MAMMOMAT 300/3000, MAMMOMAT 3000 Modular incl. Stereotactic Biopsy
Attachment and Opdimar DHHS Maintenance Instructions
(included in the Opdimar delivery)
MAMMOMAT 300/3000, MAMMOMAT 3000 Modular incl. Stereotactic Biopsy
Attachment and Opdimar Measurement Certificates
(included in the Opdimar delivery)
MAMMOMAT Novation
DR
- Opdimar Planning Guide
MAMMOMAT Novation
DR
- Opdimar Installation and Start-Up
MAMMOMAT Novation
DR
- Opdimar Wiring Diagram
MAMMOMAT 3000 Installation and Setting Instructions Update Kit for Stereotactic
Biopsy Attachment (where applicable)
Meters and appliances required 1
Protective ground wire tester (44 15 899 RV090)
Service PC (e.g. Siemens-Nixdorf PCD 3NSX/20 or similar) with connecting cable, PC to
generator (part No. 99 00 440 RE999)
Stereo calibration phantom (included in the delivery) (part No.64 30 701)
AEC calibration plexiglass, four plates measuring 150 mm x 150 mm x 19 mm and one
plate measuring 150 mm x 150 mm x 9,7 mm, part No. 65 61 232 and 65 61 224
respectively
Resolution phantom with at least 10 line pairs per mm
Tools required 1
Standard installation tools
Electric drill with 5 mm drill
Additional installation material required 1
Extension mains cord with at least three outlets, of protective earth type
Start-up, DHHS and X-ray decree area of application 1
Notes on start-up 1
Opdimar is adjusted, programmed and tested in the factory, leaving only the connection
to the on-site mains voltage, the functional tests and customer dependent adjustments to
be performed.