Power up the MAMMOMAT Stand 3 - 3
Siemens AG SPB7-250.815.02 Page 3 of 6 MAMMOMAT
Novation
DR
Medical Solutions Rev. 02 09.04 CS SD 24 System
Connections: 3
1. Connect the incoming mains cable to the mains terminal (Fig. 3) at the bottom of
the back of the stand.
2. Strain-relieve the incoming mains cable.
Power
system
Type of
connection
Power voltage Power
connections
Single phase Phase to neutral 208 V-277 V Connect to 3 and 4 (see 1/Fig. 3)
Two phase Phase to phase 208 V-400 V
phase to phase.
Note: Max 277 V
phase to neutral/
ground allowed.
Connect to 3 and 4 (see 1/Fig.
3).
Use two phases (of three) in a
normal power distribution sys-
tem.The left over wires, neutral
and the third phase may be con-
nected to terminal 1 and 2. See
Fig. 3.
Fig. 3Incoming mains cable