SPB7-250.815.02 Page 4 of 6 Siemens AG
System Rev. 02 09.04 CS SD 24 Medical Solutions
4 - 4 Detector installation
Mounting the detector 4
1. Remove the micro switches and their holders S2 and S3.
2. Remove the cover from the detector cable connector.
3. Loosen the grid holder, 2 screws (4/Fig. 6).
4. Attach the detector with the four bolts (1/Fig. 6). Make sure that the bolts are
tightened.
5. Insert the detector cable underneath the grid holder and tighten the grid holder
screws again (4/Fig. 6).
6. Connect the detector cable (3/Fig. 6) to the cable coming out of the wing (2/Fig. 6).
7. Mount the cover over the detector cable with 2 screws. Make sure that no cables
are squeezed.
Fig. 6 Detector installation
To be sure that the detector is correctly mounted and that it is
located directly below the radiation field, use the guiding lines on
the wing and make sure that the detector is in the right position
before screwing the bolts.