Adjusting the field light time 5
The illumination time of the light field can be adjusted within a certain range according to
the customer´s requirements.
1. Connect the service PC to the D750 and start the MAMMOMAT Service Software.
2. In Mainmenu select Configuration Miscellaneous Illumination time.
3. Type in the illumination time and save with <F2>.
Checking the compression device 5
1. Attach the compression plate (1/Fig. 10) if you have not already done so.
2. Check that the compression plate cannot be removed without pressing the
compression plate release button (2/Fig. 10).
3. Check the function of the knobs for manual compression/decompression
(3/Fig. 10).
4. Place a piece of soft material on the object table, for example a rolled up towel or a
piece of foam rubber (4/Fig. 10).
A shorter illumination time results in an increased life time of the
lamp.