System Rev. 02 09.04 CS SD 24 Medical Solutions
6 - 2 X-ray tube checks
Oscilloscope diagram examples
Fig. 2 Fig. 3
30 kV 1 V/T (1 T <=> 5 kV) 30 kV 1 V/T (1 T <=> 5 kV)
130 mA 1 V/T (1 T <=> 40 mA) 166 mA 1 V/T (1 T <=> 40 mA)
160 ms 20 ms/T 128 ms 20 ms/T
Fig. 4 Fig. 5
35 kV 1 V/T (1 T <=> 5 kV) 35 kV 1 V/T (1 T <=> 5 kV)
112 mA 1 V/T (1 T <=> 40 mA) 132 mA 1 V/T (1 T <=> 40 mA)
0.9 s 0.2 s/T 0.72 s 0.2 s/T
30kV, 20mAs, Mo/Mo
30kV, 20mAs, W/Rh
35kV, 100mAs, Mo/Mo
35kV, 100mAs, W/Rh