Collimator Adjustment 7 - 5
Siemens AG SPB7-250.815.02 Page 5 of 6 MAMMOMAT
Novation
DR
Medical Solutions Rev. 02 09.04 CS SD 24 System
10. Continue with Step 7 until the beam field and light field are identical.
11. Save the calibration and Exit the Service Software.
12. Record the measured values in the "Startup Protocol" on Page 11 - 1 .
Collimator Wing Difference Calibration 7
The collimator wing difference calibration has to be set to 0.
1. Start the service software on the Service PC and select:
Configuration -> Collimator -> Calibration of wing differance
2. Save the calibration and Exit the Service Software.
"
Fig. 4 Collimator wing difference calibration