MAMMOMAT
Novation
DR
SPB7-250.815.02 Page 4 of 10 Siemens AG
System Rev. 02 09.04 CS SD 24 Medical Solutions
8 - 4 Checking dose settings
Checking the AEC dose settings 8
The dose values for the AEC settings (H & D selection on the control console) have to be
checked.
As a default from the factory, the following BRICK / AEC Configuration values are set
(for further details see the Software Brick V... manual SPB7-250.816.04...):
H = (80, 70, 45)
D = (100, 80, 60)
Checking the H & D dose settings for Mo/Mo. 8
Proceed as follows:
1. Mount the Collimator Mounted Plexi phantom and the compression plate
simulator.
2. Register or create a patient and submit with Exam.
Fig. 3 Calibration - Patient register