9 - 18 Image Quality
7.
8. Set the following exposure parameter on the control console of the MAMMOMAT
Novation
DR
:
Mo/Mo, 28KV, 250mAs
9. Insert the collimator mounted plexi phantom.
10. Mount the compression plate simulator.
11. Double click on the first Image of the 8 at the Procedure: Calibration Procedure
12. Start an exposure.
Wait until the image is loaded into the viewer. The following lines are displayed in
the message bar.
Image acquisition is in progress .....
Saving processed Image .....
Loading Image ......
13. Click the Accept or Reject button.
If you want to end the Bad Pixel Mapping operation at this point,
click the "Cancel" button.
Fig. 19 Acquisition System Menu