Final procedures 10 - 3
Siemens AG SPB7-250.815.02 Page 3 of 4 MAMMOMAT
Novation
DR
Medical Solutions Rev. 02 09.04 CS SD 24 System
Service PC and measuring instruments 10
Saving the configuration file 10
Main menu:
1. Select Configuration Save config file.
2. Press <F2> to save data.
Reset the exposure counter 10
Main menu:
1. Select Service Reset exposure counter.
2. Delete the error memory with <Y>.
Deleting the error buffer 10
Main menu:
1. Select Service Delete error buffer.
2. Delete the error memory with <Y>.
Saving the MAMMOMAT Parameters 10
Main menu:
1. Select Backup Copy installation area to floppy All.
2. Leave the program with <F10> when the backup is ready.
Remove the measuring instruments 10
1. Turn OFF the MAMMOMAT.
2. Remove the service PC.
3. Remove connected measuring instruments.
Checking the protective grounding resistance 10
Using the protective ground wire tester, measure the resistance between the protective
ground terminal in the stand (Mains input terminal) and all metallic parts of the equipment
which the patient/operator may come in contact with, incl. all object tables.
The protective ground resistance must not exceed 0.1 .
Make sure that the backup floppy is inserted.
Keep the backup floppy with other documentation for the
MAMMOMAT
Novation
DR
in question.
NOTE
NOTE