MAMMOMAT
Novation
DR
SPB7-250.815.02 Page 4 of 10 Siemens AG
System Rev. 02 09.04 CS SD 24 Medical Solutions
11 - 4 Startup Protocol
Startup Settings 11
Tube high voltage, current and mAs values 11
Results from the chapter "Tube high voltage, current and mAs values" on Page 6 - 1.
kV
mAs
Anode/
filter
Measured values
Result
kV must be &5%
mAs must be &10%
30 kV
20 mAs
Mo/Mo
_____V x 5 kV =
_____V x ( _____s + _____s + _____s) x 40 mA =
______ kV
______ mAs
30 kV
20 mAs
W/Rh
_____V x 5 kV =
_____V x ( _____s + _____s + _____s) x 40 mA =
______ kV
______ mAs
35 kV
100 mAs
Mo/Mo
_____V x 5 kV =
_____V x ( _____s + _____s + _____s) x 40 mA =
______ kV
______ mAs
35 kV
100 mAs
W/Rh
_____V x 5 kV =
_____V x ( _____s + _____s + _____s) x 40 mA =
______ kV
______ mAs
25 kV
100 mAs,
Mo/Mo
_____V x 5 kV =
_____V x ( _____s + _____s + _____s) x 40 mA =
______ kV
______ mAs
25 kV
100 mAs
W/Rh
_____V x 5 kV =
_____V x ( _____s + _____s + _____s) x 40 mA =
______ kV
______ mAs
30 kV
10 mAs
Mo/Mo
_____V x 5 kV =
_____V x ( _____s + _____s + _____s) x 40 mA =
______ kV
______ mAs
Tab. 2 Tube high voltage, current and mAs values