Startup Protocol 11 - 5
Siemens AG SPB7-250.815.02 Page 5 of 10 MAMMOMAT
Novation
DR
Medical Solutions Rev. 02 09.04 CS SD 24 System
30 kV
10 mAs
W/Rh
_____V x 5 kV =
_____V x ( _____s + _____s + _____s) x 40 mA =
______ kV
______ mAs
25 kV
-
Mo/Mo
_____V x 5 kV =
_____V x ( _____s + _____s + _____s) x 40 mA =
______ kV
______ mAs
Display ______ mAs
25 kV
-
W/Rh
_____V x 5 kV =
_____V x ( _____s + _____s + _____s) x 40 mA =
______ kV
______ mAs
Display ______ mAs
kV
mAs
Anode/
filter
Measured values
Result
kV must be &5%
mAs must be &10%
Tab. 2 Tube high voltage, current and mAs values