EasyManua.ls Logo

Siemens MAMMOMAT Novation DR - Page 89

Siemens MAMMOMAT Novation DR
104 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Startup Protocol 11 - 5
Siemens AG SPB7-250.815.02 Page 5 of 10 MAMMOMAT
Novation
DR
Medical Solutions Rev. 02 09.04 CS SD 24 System
30 kV
10 mAs
W/Rh
_____V x 5 kV =
_____V x ( _____s + _____s + _____s) x 40 mA =
______ kV
______ mAs
25 kV
-
Mo/Mo
_____V x 5 kV =
_____V x ( _____s + _____s + _____s) x 40 mA =
______ kV
______ mAs
Display ______ mAs
25 kV
-
W/Rh
_____V x 5 kV =
_____V x ( _____s + _____s + _____s) x 40 mA =
______ kV
______ mAs
Display ______ mAs
kV
mAs
Anode/
filter
Measured values
Result
kV must be &5%
mAs must be &10%
Tab. 2 Tube high voltage, current and mAs values