10 Important Notes
MAMMOMAT Novation DR SPB7-250.815.06.02.02 Siemens AG
12.05 CS SD 24
Page 10 of 116
Medical Solutions
Factory Adjustments 0
The MAMMOMAT Novation
DR
is adjusted, programmed and tested in the factory, leav-
ing the adaptation to the on-site mains voltage, adjustment of the AEC and the functional
tests to be performed.
When the measurements to be made (kV, mAs, etc.) are within the tolerances stated in
these instructions, this confirms that the settings made in the factory have not changed
and the equipment is fully serviceable.
The equipment is tested with 400 V, 2-phase in the factory.