20 Power up the MAMMOMAT
MAMMOMAT Novation DR SPB7-250.815.06.02.02 Siemens AG
12.05 CS SD 24
Page 20 of 116
Medical Solutions
Connections 0
Fig. 5: Three incoming mains cables
1. Connect the incoming mains cable to the mains terminal ((Fig.5/p.20)) at the bottom
of the back of the stand.
2. Strain-relieve the incoming mains cable.
Power system Type of connection Power voltage Power connections
Single phase Phase to neutral 208 V-277 V Connect to 3 and 4 (see No.1
/(Fig.5/p.20))
Two phase Phase to phase 208 V-400 V phase to
phase.
Note: Max. 277 V
phase to neu-
tral/ground allowed.
Connect to 3 and 4 (see No.1
/(Fig.5/p.20))
Use two phases (of three) in a nor-
mal power distribution sys-
tem.The leftover wires, neutral
and the third phase may be con-
nected to terminal 1 and 2. See
(Fig.5/p.20).