EasyManua.ls Logo

Siemens MAMMOMAT Novation DR - Adjusting the Switches

Siemens MAMMOMAT Novation DR
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
30 Detector installation
MAMMOMAT Novation DR SPB7-250.815.06.02.02 Siemens AG
12.05 CS SD 24
Page 30 of 116
Medical Solutions
Adjusting the switches 0
Fig. 13: Micro switch installation
1. Mount the micro switches S2 and S3 (Fig. 13 / p. 30) along the markings.
2. Adjustments of S2 move the grid to the front. S2 must switch 2 mm (+1 mm tolerance)
before the grid reaches the end position.
3. Adjustment of S3 Move the grid to the back and insert the magnification table. S3 must
switch 2 mm (+1 mm tolerance) before the magnification table reaches the end position.
NOTE
The magnification table is optional.
When no magnification is supplied with the system, the micro
switch has to be mounted along the markings.
Remember to check the switch adjustment if the customer orders
a magnification table.

Table of Contents

Other manuals for Siemens MAMMOMAT Novation DR

Related product manuals