62 Checking dose settings
MAMMOMAT Novation DR SPB7-250.815.06.02.02 Siemens AG
12.05 CS SD 24
Page 62 of 116
Medical Solutions
4. Click Modify settings.
Fig. 40: BRICK_Modify settings
5. To adjust the
- H value,
go to Modify low dose scaling (The default values are 80, 70, 45) and enter the new
values, separated by a comma, e.g. 80,70,45 in the Low Dose Scaling Value field and
click Change Low Dose Scaling to save it.
- D value,
go to Modify high dose scaling (The default values are 100, 80, 60) and enter the new
values, separated by a comma, e.g. 100,80,60 in the High Dose Scaling Value field
and click Change High Dose Scaling to save it.
6. Proceed from where you were originally referenced.