EasyManua.ls Logo

Siemens MAMMOMAT Novation DR - Creation of a New Pixel Map (if Necessary)

Siemens MAMMOMAT Novation DR
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
76 Image Quality
MAMMOMAT Novation DR SPB7-250.815.06.02.02 Siemens AG
12.05 CS SD 24
Page 76 of 116
Medical Solutions
12. Wait approx. 10 seconds for the image to be calculated; observe the status bar.
Once the Accepted images counter increases, the image has been successfully
accepted.
13. Double-click the next image in the Procedure: Calibration and proceed with step 10
until you have a total of 8 accepted images.
14. Select Examination -> End Examination from the syngo menu bar to exit the calibra-
tion.
Creation of a new pixel map (if necessary) 0
During the calibration function, you have the option of creating a new pixel map of the
detector. The pixel map indicates which pixels in the detector‘s array might not be reading
X-ray densities correctly and therefore should be ignored when the image is acquired.
This is similar to a bad spot map for a computer‘s hard disk.
A new pixel map should be created when an excessive amount of pixel dropout is occur-
ring repeatedly on images displayed in the image Preview window.
To create a new pixel map:
1. Start the service software from the syngo menu bar:
Options -> Service -> Local Service
NOTE
The Service Level 5 or higher is necessary to do the bad pixel map-
ping.
2. Select Acquisition System.
Fig. 55: Service Home Page

Table of Contents

Other manuals for Siemens MAMMOMAT Novation DR

Related product manuals