EasyManua.ls Logo

Siemens MAMMOMAT Novation DR - Disabling the Film Exposure

Siemens MAMMOMAT Novation DR
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Siemens AG SPB7-250.815.06.02.02 MAMMOMAT Novation DR
12.05 CS SD 24
Final procedures 89
Page 89 of 116
Medical Solutions
11-
10Final procedures
Disabling the film exposure 0
The film exposure switch has to be disabled.
1. Connect the service PC to the MAMMOMAT stand and start the service software.
2. Go to:
Configuration -> Miscellaneous -> Allow film exposure
and make sure that the Switch is set to OFF.
Fig. 60: Set Allow film exposure to OFF