Siemens AG SPB7-250.844.01 Page 9 of 10 MAMMOMAT
Novation
DR
Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E 4-9
GENERATOR: LIST OF SWITCHES,
POTENTIOMETERS, TEST POINTS, LEDs
0
Switches: 0
Potentiometers 0
Test points 0
Switch Page Module Description
S700 4-5 Mains input
(Filter unit)
On/Off circuit breaker for mains input, inside the covers. The
breaker is intended for service/maintenance as an extra possibility
to turn off the mains for safety at work with the unit.
It shall be in On position when the covers are mounted. The breaker
also protects the generator, it trips at overcurrent (30A) plus has a
trip coil used for DC-bus overvoltage protection.
Potentiometer Page Module Description
Bias Mo 5-6 Tube power
supply
Adjustment of bias voltage for Molybden (Mo).
Bias W 5-6 Tube power
supply
Adjustment of bias voltage for Tungsten (W).
Anode drive freq. 5-7 Tube power
supply
Adjustment of drive freq. 3 - 5 Hz from its nominal value, for
decreasing noise that can occur due to mechanically vibrations
from the tube assembly.
Test point Page Module Description
0VA 5-6 Tube power
supply
Logic supply voltage: 0 volt analog
Note: 0VA = 0V_sec
FIL_CURR_READ 5-6 Tube power
supply
Analog signal proportional to the true filament current.
The relation between the filament current and the analog signal is
1A/V.
MA_READ 5-6 Tube power
supply
Analog signal proportional to the true tube current.
The relation between the tube current and the analog signal is
40mA/V.
I_OUT 5-8 HV inverter Analog signal proportional to the output inverter current fed to the
HV tank.
HV_READ 5-8 HV inverter Analog signal proportional to the true tube voltage.
The relation between the tube voltage and the analog signal is
5kV/V.
HV_WISH 5-8 HV inverter Analog control signal for setting the magnitude of the tube voltage
(high voltage).
The relation between the tube voltage and the analog control signal
is 5kV/V.
0VA 5-6 HV inverter Logic supply voltage: 0 volt analog
Note: 0VA = 0V_sec