1-8 February 5, 2004 FFDM-S Service Manual
CAUTION
Some of the electronic components used in this product may be damaged
by static electricity or electrostatic discharge. Unpack and handle using appropriate
precautions.
CAUTION
Dielectric strength test caution: (Statement from Condor DC Supplies Inc.
regarding their GSM28-5 power supply) "When performing Dielectric Strength Tests,
catastrophic failure of the unit may result if a Dielectric Strength test voltage greater
than 1800 Vac is applied between primary and secondary circuits. The components
providing isolation from primary to secondary cannot be tested while installed in the
power supply without overstressing basic (primary to ground) insulation. All isolating
components are individually 100% tested at 4800 Vac prior to installation."