5-8 February 5, 2004 FFDM-S Service Manual
Upgrading Detector Software Components
NOTE: Typically, you will be upgrading an Detector software component only at the direction of Direct
Radiography Corp., or if you suspect that the current software component is corrupt.
To upgrade one of the Detector’s software components:
1. Click the Software Upgrade link on the Array Control Panel page, or the
Upgrade link on any of the other Array pages.
The Upgrade Detector Subsystem (Array) Software page displays.
NOTE: For security purposes, you may be prompted to enter a login name and password before
executing this function. For more information, refer to “Function Security Levels and Login Prompts”
on page 5-2.
2. Select the software component that you want to upgrade.
3. Click the Browse button, then navigate to and select the appropriate software
component file from your Acquisition Workstation’s file system.
The file you selected is displayed in the Please select the file to transfer
field
CAUTION
The PXCM Web Tool does not check that the appropriate file type has been
selected. If you select the wrong file to upload, you could cause the PXCM to stop
operating.