PXCM Web Tool February 5, 2004 5-21
4. Click the Browse button, then navigate to and select the appropriate PXCM
software file from your Acquisition Workstation’s file system.
The file you selected is displayed in the Please select the file to transfer
field.
5. For the Select Boot Mode option, select normal, unless directed to select a
different mode by Direct Radiography Corp. Support.
6. Select the Sync Safe & Normal Modes? option.
NOTE: To clear your entries or selections, click the Reset Form button.
7. When you are ready to start the software file transfer, click the Upgrade
button.
The file transfer status is displayed on the page.
8. After the transfer is complete, reboot the PXCM to have it start up using the
software you just transferred. For information about rebooting the PXCM,
refer to “Powering the Detector On and Off” on page 5-9.
Rebooting the PXCM
To reboot the PXCM, click the Reboot the Brick link on the Array Control Panel
page, or the Reboot link on any of the other PXCM pages. You will be prompted to
confirm the reboot operation.
NOTE: For security purposes, you may be prompted to enter a login name and password before
executing this function. For more information, refer to “Function Security Levels and Login Prompts”
on page 5-2.
CAUTION
The PXCM Web Tool does not check that the appropriate file type has been
selected. If you select the wrong file to upload, you could cause the PXCM to stop
operating.