EasyManua.ls Logo

Siemens MAMMOMAT NovationDR - Interpreting the PXCM Status LEDs; Reporting DirectRay Log Information

Siemens MAMMOMAT NovationDR
95 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-2 February 5, 2004 FFDM-S Service Manual
Interpreting the PXCM Status LEDs
The PXCM includes three status LEDs (shown in Figure 7-1 below):
V1—Fiber-optic communications status. A red LED that lights to indicate
lack of communication from the FFDM-S. Likely causes are misconnected fiber
optic lines or the Detector being turned off.
V2—PXCM status. A green LED that flashes to indicate normal operation of
the PXCM. If this LED does not light for a significant amount of time (that is,
2 to 3 minutes), reboot the PXCM.
V3—Power status. A green LED that lights to indicate that the PXCM is
providing power to the Detector.
Figure 7-1.
DirectRay PXCM Status LEDs Location
Reporting DirectRay Log Information
You may be asked by your central service support group to submit DirectRay log
information related to a problem. You use the PXCM Web Tool, described in
Chapter 5, to gather the relevant log files. You can then send the file to your central
service support group for problem diagnosis.
GS20_418.cdr
PXCM Status LEDs

Table of Contents

Other manuals for Siemens MAMMOMAT NovationDR

Related product manuals