BW500 and BW500/L
Operating Instructions, 12/2016, A5E33482052-AD
153
f
indicates factory set value
Locks out the program editor so that parameter values for P001 through P999 cannot be
changed. This however, does not prevent the access to the parameters for viewing.
Programming is locked out if the value of P000 is other than 1954.
1954 = unlocked
f
This is the minimum parameter programming required before attempting a calibration and
successful entry into the
mode.
Selects the language for communication with the BW500 and BW500/L
P002 Test Reference Selection
Selects the type of test reference used to represent a material load: weight, chain, or electronic.
the weight that is supplied specific to the scale
optional, sized specific to scale and conveyor
electronic: calibration based on automatic calculation of the
mV span from the load cells
1 = weight
f
2 = chain
3 = ECal