Parameters
12.5 Online Calibration Options (P355 - P358)
BW500 and BW500/L
168 Operating Instructions, 12/2016, A5E33482052-AD
P350 Calibration Security
P350 Calibration Security
entry: 0 = no additional
f
Yes Yes Yes
1 = in addition to
P000 lock; no
Yes No Yes
2 = in addition to
P000; no zero, no
No No Yes
3 = in addition to
P000; no zero, no
span, no totalizer
No No No
Online Calibration Options (P355 - P358)
Online calibration options must be enabled (P355 = 1) before they become available.
Not available with BW500/L.
P355 Online Calibration Feature
Enables Online Calibration.
0 = OFF
f
1 = ON
P356 Online Calibration Reference Weight
Enter the weigh bin reference weight, (in units selected in P005), range 0.000 to 99999 (
f
=
0.000).