Parameters
12.10 ECal Parameters (P693 P698)
BW500 and BW500/L
186 Operating Instructions, 12/2016, A5E33482052-AD
Enter the inclination in degrees.
Upon entry, display jumps to P017 to display the design load value as the test load.
12.10.7
P699 ECal mV Span
Enter the mV span corresponding to 0 to 100% full scale load.
Upon entry, display jumps to P017 to display the design load value as the test load.
12.10.8
P735 Back Light
Sets the intensity of the back light for the LCD (
f
= 10).
Entry:
0 = off
1 to 10 = low to high
12.10.9
P739 Time Zone
The offset from Greenwich mean time (GMT) of local time.
This parameter does not affect any timed events because all times are local. It can be accessed
by a remote computer for synchronization purposes.
Enter local time zone 12 to 12