Start Up
6.3 Initial Start Up
BW500 and BW500/L
Operating Instructions, 12/2016, A5E33482052-AD
51
To reset a parameter value
Press the following key:
P011 Design Rate:
E
Press the following key:
P011 Design Rate:
V
Value is reset to factory value. For example: 0.00 kg/h
To operate BW500 and BW500/L in the
mode, the unit must undergo an initial
programming to set up the base operating parameters.
Attempting to enter the
mode without satisfying the program requirements forces the
program routine to the first missing item.
Initial start up of BW500 and BW500/L consists of several stages, and assumes that the physical
and electrical installation of the belt scale and speed sensor, if used, is complete:
● power up
● programming
● load cell balancing
● zero and span calibration