Start Up
6.4 Load Cell Balancing
BW500 and BW500/L
Operating Instructions, 12/2016, A5E33482052-AD
63
6.4.3
Zero Calibration
Note
To obtain an accurate and successful calibration, ensure that the required criteria are met.
Refer to
Calibration Criteria (Page 199).
Press the following key:
Zero Calibration: Current Zero
Clear belt. Press ENTER to Start
0
The current zero count
Press the following key:
Initial Zero Calibration. In progress
Current Reading:
%
#####
The zero count being calculated while calibration is in progress
The duration of the Zero calibration is dependent on speed (P014), length (P016), and
revolutions (P360) of the belt.
Calibration Complete. Deviation
Press ENTER to accept value:
0.00
551205
The deviation from previous zero. For an initial zero there is no previous zero; hence the deviation is
0. For example: the new zero count, if accepted
Press the following key:
Zero Calibration. Current Zero
Clear belt. Press ENTER to Start
551205
Accepting the Zero returns to start of Zero. A new Zero can be performed, or continue to Span.