Recalibration
7.2 Material Tests
BW500 and BW500/L
70 Operating Instructions, 12/2016, A5E33482052-AD
Access P019 and enter EDIT mode
P019 Manual Span Adjust
Select 1-% Change 2-Material Test
E
Press the following keys:
P598 Span Adjust Percentage
Enter Calculated +/- error
V
Press the following key:
P598 Span Adjust Percentage
Enter Calculated +/- error
E
Press the following keys:
if % change is negative, remember to enter the minus sign, for example, -1.3.
P017 Test Load Weight:MS1
V
For example: the new test load value is displayed
Material Test
Access P019 and enter EDIT mode
P019 Manual Span Adjust
Select 1-% Change 2-Material Test
E
Press the following keys: