Recalibration
7.4 Recalibration
BW500 and BW500/L
Operating Instructions, 12/2016, A5E33482052-AD
75
Press
Zero Calibration. Current Zero
Clear belt. Press ENTER to start
551413
For example, the current zero count.
Note
This is the end of zero calibration. Proceed with span recalibration or return to RUN
.
Use direct zero entry (P367) when replacing software or hardware, if it is not convenient to
perform an initial zero. A record of the last valid zero count is required.
Access P367 and enter EDIT mode
P367 Direct Zero Entry
E
Press the following keys in a sequence:
Zero Calibration. Current Zero
V
For example, the last valid zero count.