10 General
MOBILETT XP Digital SPR8-230.814.30.03.02 Siemens AG
11.05 CS SD 24
Page 10 of 70
Medical Solutions
DANGER
To avoid electrical shock from components under voltage, also be
aware that:
The capacitors of the capacitor bank can be electrically charged
even when the system is switched off and the power cable is dis-
connected!
Individual capacitors of the capacitor bank may have been discon-
nected from the discharging circuit, erroneously or as a result of
failure, and can still be electrically charged!
Disconnect the battery blocks in the XP Digital prior to mainte-
nance and service work!
Carelessness can result in death or serious bodily injury.
¹ Make sure that no parts or tools fall into the unit;
¹ Do not touch potentially dangerous
components(Fig.2/p.11)
¹ If loose parts must be removed from the unit, use only
insulated tools;
¹ Protect the work area so that no other persons are able
touch the unit while the covers are open or removed!
Switch the unit off before servicing or maintenance. Always disconnect the power plug
first.
Make sure that the main switch is switched off.
The capacitor bank discharges to < 40 V in approx. 15 minutes.
¹ The safety covers can be removed after this period has elapsed.
Prior to performing any work, it must be verified that areas with dangerous voltage are
voltage-free (Fig.2/p.11).