10 General
MOBILETT XP Digital SPR8-230.814.30.03.02 Siemens AG
11.05 CS SD 24
Page 10 of 70
Medical Solutions
DANGER
To avoid electrical shock from components under voltage, also be
aware that:
The capacitors of the capacitor bank can be electrically charged
even when the system is switched off and the power cable is dis-
connected!
Individual capacitors of the capacitor bank may have been discon-
nected from the discharging circuit, erroneously or as a result of
failure, and can still be electrically charged!
Disconnect the battery blocks in the XP Digital prior to mainte-
nance and service work!
Carelessness can result in death or serious bodily injury.
¹ Make sure that no parts or tools fall into the unit;
¹ Do not touch potentially dangerous
components(Fig.2/p.11)
¹ If loose parts must be removed from the unit, use only
insulated tools;
¹ Protect the work area so that no other persons are able
touch the unit while the covers are open or removed!
• Switch the unit off before servicing or maintenance. Always disconnect the power plug
first.
• Make sure that the main switch is switched off.
• The capacitor bank discharges to < 40 V in approx. 15 minutes.
¹ The safety covers can be removed after this period has elapsed.
• Prior to performing any work, it must be verified that areas with dangerous voltage are
voltage-free (Fig.2/p.11).