Siemens AG SPR8-230.831.30.03.02 MOBILETT XP Digital
11.05 CS SD 24
General maintenance information 23
Page 23 of 74
Medical Solutions
The values determined, including the measuring points, must be recorded and assessed
in the protective conductor resistance report.
The measuring procedure and the measuring device used (designation and serial num-
ber) must also be documented.
NOTE
File the protective conductor resistance report in the "Certificates"
register no. 9 in the system binder.
NOTE
A new report must be created if the protective conductor resis-
tance measurements are not documented.
A new protective conductor resistance report can be found in the
startup instructions SPR8-230.814.30.03. Separate this report from
these instructions, fill it out, and file it in the "Certificates" register
no. 9 in the system binder. If values are newly determined, they
must be recorded as first measured values. The evaluation is omit-
ted in this case.