General maintenance information 25
Page 25 of 74
Medical Solutions
However, please note the minimum resolution of the leakage current measuring device
and any additional manufacturer information restricting the use of the measuring device.
Fig. 2: Measuring circuit for measuring the system leakage current according to the differential
current method in compliance with DIN VDE 0751-1/2001-10, Fig. C6 for protection
class I.
Pos. 1 = System
Pos. 2 = Application part type B (if available)
Pos. 3 = Measurement setup (integrated into measuring device)
If the direct measurement of the leakage current is used (measurement setup according
to (Fig.3/p.26)), the system must be insulated during the measurement and must not be
touched.